Comparison of Fracture Torque and Level of Fracture of Two Orthodontic Mini-Screw (An In Vitro Study)

Sabah Mohammed Jamil, Omar and Himdad Majeed, Bahoz (2023) Comparison of Fracture Torque and Level of Fracture of Two Orthodontic Mini-Screw (An In Vitro Study). Eurasian Journal of Science and Engineering, 9: 2. pp. 241-251. ISSN 24145629

[thumbnail of Research Article] Text (Research Article)
Article_EJSE_25-06-2023.pdf - Published Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.

Download (1MB)

Abstract

Certain malocclusions may benefit from the use of mini-screw anchoring control during orthodontic treatment. In contrast, the failure of little screws will have a substantial influence on treatment efficiency and effectiveness. Twenty orthodontic mini-implants of (Tomas. Dentaurm. Germany; GNI, Korean) were inspected in an in-vitro study. A digital torque gauge was positioned perpendicular to the bone surface and placed into the 10 mm thick bovine femoral cortical bone in the same manner as mini-implants. The insertion done through the hand driver simulations of insertion intraorally. The Tomas offered a high fracture torque value compared with GNI. GNI gave a high percentage of fracture at an apical level while Tomas gave at the middle of it.

Item Type: Article
Uncontrolled Keywords: Mini Screws; Anchorage; GNI Mini Screws; Tomas Mini Screws.
Subjects: R Medicine > RK Dentistry
Divisions: Department of Dentistry > Research papers
Depositing User: ePrints Depositor
Date Deposited: 06 Aug 2025 06:29
Last Modified: 06 Aug 2025 06:29
URI: https://eprints.cihanuniversity.edu.iq/id/eprint/4469

Actions (login required)

View Item
View Item